SPECIFICATIONS
SpectraDynamics, Inc. 1849 Cherry Street Unit 2, Louisville CO, 80027 Phone (303) 665 1852 Fax (303) 604 6088
www.spectradynamics.com
48
PARAMETER CONDITIONS TYP UNITS
Frequency stability s
y
(t)
Ref. 5 MHz, +13 dBm
Pair measurement
averaging time
1 s
10 s
100 s
1000 s
100000 s
1000000 s
1·10
-13
1·10
-14
1·10
-15
2·10
-16
2·10
-17
2·10
-17
-
Phase noise L(f)
Carrier 6.834 GHz
Pair measurement
offset frequency
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
-82
-94
-112
-120
-120
-130
dBc/Hz
Phase noise L(f)
Carrier 100 MHz
offset frequency
10 Hz
100 Hz
1 kHz
>10 kHz
-122
-132
-159
-174
dBc/Hz
Phase noise L(f)
Carrier 10 MHz
offset frequency
1 Hz
10 Hz
100 Hz
>1 kHz
-112
-142
-160
-167
dBc/Hz
Phase noise L(f)
Carrier 5 MHz
offset frequency
1 Hz
10 Hz
100 Hz
>1 kHz
-118
-150
-167
-173
dBc/Hz
Temperature coefficient
@ 6.834 GHz
0.5 ps/K
External reference
5.0 MHz ± 2.0·10
-8
+7 dBm to +15 dBm
Rack-mount enclosure
Size: 3.5” X 16” X 19”
Weight: 20 lbs